Status Registers
Use Status Registers to Determine the State of Analyzer Events and Conditions
STATus:QUEStionable:INTegrity:UNCalibrated
Condition Register Bits
Decimal
Bit
Value
0
0
1
2
2–14
–––
15
32768
2-26
Description
Oversweep (Meas Uncal) A 1 in this position
indicates that the anlyzer is in a state that could lead
to uncalibrated measurements. This is typically
caused by sweeping too fast for the current
combination of span, resolution bandwidth, and video
bandwidth. Auto coupling may resolve this problem.
Signal Ident ON A 1 in this bit position indicates
that amplitude measurements may be in error due to
signal identification routines being active. Amplitude
accuracy is degraded when signal identification is
active. (Agilent ESA model E4407B Option AYZ,
External Mixing, only).
Reserved These bits are not used by the analyzer, but
are for future use with other Agilent products.
Always Zero (0) This bit is always set to 0.