Troubleshooting
Error Code Tables
Table 4.3
Error Lane Error Codes (Continued)
Sensor
SN081 OFF
SN082 OFF

Bulk Inlet Error Codes

Table 4.4
Bulk Inlet Error Codes
Sensor
AD-1 ERR
AD-2 ERR
DM013A ON
DM013B ON
DM013 OFF
DM015A ON
DM015B ON
DM015 OFF
DM018A ON
DM018B ON
DM018 OFF
4-18
Problem
Tube size detection error.
Sensor did not turn off
correctly with no sample
tube present. Tube size
detection of the sample
tube at the BCR position
may be incorrect.
Problem
AD converter for tilt sensor
fails to initialize.
Loading Position 1 Drop-
Guide is not working
properly. *B ON Error;
Releasing motion failed. *A
ON Error; Guide motion
failed. *OFF Error; Sensors
at both ends became ON.
Loading Position tube-
pusher is not working
properly. *B ON Error;
Releasing motion failed. *A
ON Error; Guide motion
failed. *OFF Error; two of
three sensors (MS015,
MS016 or MS017) became
ON simultaneously.
Tube Z-axis is not working
properly. *B ON Error;
Down motion failed. *A ON
Error; Upward motion
failed. *OFF Error; two of
three sensors (level
(DM018), Cap present
(PS018 or PS019) became
ON simultaneously.
1.
Restart
Select
on the Control Panel to verify sensor
check works correctly.
2.
Inspect for obstruction on tube size detection arms.
3.
Clear
Select
on the Control Panel to proceed to the
next sample tube.
Turn off the line and restart.
1.
Confirm that the sensors are working correctly.
2.
Confirm that the tube is positioned correctly and that there
are no obstructions.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for any obstructions, wrinkled labels, or mishandled
tubes. If necessary, remove any tubes, replace labels, or
reposition tube in the sample holder.
2.
Confirm that the sensors are working correctly.
3.
Confirm that the air compressor is on and the auto switch
is working correctly.
4.
Confirm that the compressed air supply is within the
normal pressure range.
5.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sensors are working correctly.
2.
Confirm that the tube is seated correctly.
3.
Restart
Select
on the control panel to resume
operation.
Solution
Solution
B31451AE