Table 4.8
T Lane Error Codes
Sensor
BUZZ01-0*
BUZZ02-0*
BUZZ03-0*
4-40
Problem
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
1.
Restart
Select
on the control panel to retry.
2.
Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3.
If the error persists, select
to release the carrier.
Note
#1: Where bar code readers and RFID readers
coincide, the sample tube must be removed from
the carrier before the Clear option is available. If
the bar code label cannot be read, remove the
tube and select Clear
#2: The BUZZx-0y refers to which BCR (XX) is
unable to determine how to handle the sample
tube and why it is unable to handle the sample
tube(0y).
-01 indicates the sample tube bar code label is unreadable.
Repeat the read operation or reprint the bar code label and re-
apply it to the tube and repeat the read operation.
-02 indicates the sample tube bar code label was read, however
the ID indicates this tube should be somewhere else on the
system. This is frequently due to a mis-read of an aliquot tube
either here or somewhere else. The sample tube will need to be
removed and select Clear
-03 indicates the sample tube bar code label was read, however
the ID indicates the tube should not be on the system at all.
This is frequently due to a secondary label which may be
partially in view. Correct any secondary label and repeat, and if
the problem continues, remove the sample tube and select
Clear
.
Solution
Clear
on the control panel
.
to recover from the error.