Beckman Coulter Power Express Instructions For Use Manual page 96

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.4
Bulk Inlet Error Codes (Continued)
Sensor
DM111A ON
DM111B ON
DM111 OFF
DM115A ON
DM115B ON
DM115 OFF
DM121A ON
DM121B ON
DM121 OFF
DM123A ON
DM123B ON
DM123 OFF
DM125A ON
DM125B ON
DM125 OFF
DM127A ON
DM127B ON
DM127 OFF
4-20
Problem
Tube Supply Box Z-axis
Lifter is not working
properly. *A ON Error;
failed to close. *B ON Error;
failed to open. *OFF Error;
Sensors on both ends
became ON.
Tube Supply Box door is not
working properly. *A ON
Error; Upward motion
failed. *B ON Error;
Downward motion failed.
*OFF Error; Sensors on both
ends became ON.
Right side tube transfer belt
stopper 1 is not working
properly. *A ON Error;
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
Right side tube transfer belt
stopper 2 area is not
working properly. *A ON
Error; Failed to move to the
supply box side. *B ON
Error; Failed to move to the
tube clamp holds side. *OFF
Error; Sensors on both ends
became ON.
Right side tube transfer belt
stopper 3 is not working
properly. *A ON Error;
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
Right side tube transfer belt
stopper 4 is not working
properly. *A ON Error;
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
1.
Confirm that the sensors are working correctly.
2.
Inspect for and remove any obstructions.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Restart
Select
on the control panel to resume
operation.
Solution