Table 4.3
Error Lane Error Codes (Continued)
Sensor
Error at the Y-axis cylinder
SL021A ON
on the carrier transfer arm.
SL021B ON
ON error: The cylinder does
SL021 OFF
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Error at the Z-axis cylinder
SL022A ON
on the carrier transfer arm.
SL022B ON
ON error: The cylinder does
SL022 OFF
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
SL025 ON
is not working correctly.
SL025 OFF
ON error: The cylinder does
SL026 ON
not respond correctly to an
SL026 OFF
ON signal from a PLC.
SL027 ON
SL027 OFF
OFF error: The cylinder
SL028 ON
does not respond correctly
to an OFF signal from a PLC.
SL028 OFF
Carrier guide is not working
SL029 ON
correctly.
SL029 OFF
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
B31451AE
Problem
1.
Confirm that the air compressor is on and the auto switch
is working correctly.
2.
Confirm that the compressed air supply is within the
normal pressure range.
3.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-15