Selecting Remap Destination At The Inlet Control Panel; Priority - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Inlet and Outlet Module Procedures
Load Racks into the Inlet Module
Figure 2.2 Inlet Module Control Panel Rack Legend
Table 2.2
E
ST
RT
BP
RM
RM-A
RM-R
6
If you place a rack in the Inlet that you do not want processed, either remove it or select
the corresponding rack position so that the icon reverts to
7
Close the Inlet rack drawers.
8
Select Restart
Selecting Remap Destination at the Inlet Control Panel (V3.0 and Later)
When placing a Remap rack in the Inlet, the system selects Remap Refrigerated Storage
(RM-R) as the remap destination. This is the default destination for all remaps. All
remapped samples of all sample types sort to the refrigerated Storage modules. By
selecting the remap rack position on the Control Panel, you can change this from RM-R to
2-4
Rack Type Abbreviations
Abbreviation
Caution
Confirm that the control panel legend matches the rack types in the rack drawer.
If the rack types do not match, repeat the rack change. If the rack type mismatch
continues, contact your Beckman Coulter representative.
.
Definition
Empty

Priority

Routine
Centrifuge Bypass
Remap
Remap Ambient Storage (V3.0 and later)
Remap Refrigerated Storage (V3.0 and later)
.