Inlet Error Codes - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.1
Code
OFF
ON
OP CALL
PAS
PM
POS
PR
PRINT SYNTX
PRNT CUT
PRNT HEAT
PRNT PAUSE
PS
REG RST
RF
RF INI
RST
SL
SN
TIP ERR
Z# Shock

Inlet Error Codes

Table 4.2
Inlet Error Codes
Sensor
BUZZ01-0*
BUZZ02- 0*
4-2
Error Code Definitions (Continued)
1.
A cylinder is in an abnormal condition with an OFF status.
2.
A sensor detects an abnormal condition with an OFF status.
A cylinder is in an abnormal condition with an ON status.
A sensor detects an abnormal condition with an ON status.
Operation Call (contact your Beckman Coulter Field Service Engineer )
Pass
Pulse Motor
Positioning
Label Printer
Printer Syntax
Printer Cut
Printer Heat
Printer Pause
Photo Sensor
Register Reset
RFID
RF Initialization
Reset
Cylinder
Sensor
Tip Error
Shock Sensor (Z-axis)
Problem
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
Definition
1.
Restart
Select
on the control panel to retry.
2.
Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3.
If the error persists, select
to release the carrier.
Solution
Clear
on the control panel
B31451AE