Table 4.3
Error Lane Error Codes (Continued)
Sensor
Lane 1 Camera fails to
CA1UNT ERR
initialize.
CA1SND ERR
CA1RCV ERR
CA1COM ERR
CA1CMD ERR
CA1CON ERR
CA1PNG TIM
CA1PNG ERR
Lane 2 Camera fails to
CA2UNT ERR
initialize.
CA2SND ERR
CA2RCV ERR
CA2COM ERR
CA2CMD ERR
CA2CON ERR
CA2PNG TIM
CA2PNG ERR
Passage sensor error. A
FS011 PAS
passage sensor does not
PS025 PAS
detect a carrier within a
predetermined timeout
period.
A carrier is under the carrier
PS025 ON
transfer arm.
RFID initialization error.
RF01 INI
RF02 INI
B31451AE
Problem
1.
Select
2.
If problem still continues, power off track to reset the
camera unit.
3.
If the error still occurs, disable lane 1 operation.
4.
If the error still occurs, contact your Beckman Coulter
representative.
1.
Select
2.
If problem still continues, power off track to reset the
camera unit.
3.
If the error still occurs, disable lane 2 operation.
4.
If the error still occurs, contact your Beckman Coulter
representative.
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
3.
If no sample tube carrier is there or stopped, the passage
of the carrier did not trigger the sensor. Use a metal object
to trigger the sensor manually. Select
1.
Inspect to see if there is a carrier under the carrier transfer
arm. If found, remove the carrier from the area.
2.
Confirm that the sensor is working correctly.
1.
Confirm that the RFID cable is connected correctly.
2.
Reboot the module.
3.
Place the module in
Solution
Restart
to repeat the operation.
Restart
to repeat the operation.
Restart
on the control panel to resume
Run
mode to resume operation.
4
Troubleshooting
Error Code Tables
Restart
. I
4-13