Beckman Coulter Power Express Instructions For Use Manual page 29

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Figure 1.6 Inlet Universal Racks
1. Universal Priority Rack: RED
2. Universal Routine Rack: BLUE
Each rack is keyed to identify the rack type to the system, and to fit in the Inlet rack drawer
in only one direction.
If you place the rack in the drawer backwards (position 50 facing the back of the Inlet), the
system will not sense the rack, and the drawer might not close completely, delaying sample
processing.
Rack positions in the Inlet rack drawer are numbered as shown in the following table.
Dynamic Inlet Error Lane
After the gripper arm places sample tubes into sample tube carriers and detects tube
dimensions, the bar code reader scans each sample tube bar code label, and the system
verifies the associated sample programming. The system routes sample tubes to the Error
Lane (refer to
• The system did not load the sample tube from a rack or Bulk Inlet Module.
• The system cannot read the bar code.
• The bar code is the wrong length or type.
• There is no sample programming for the sample ID, or there is a programming conflict.
• You tried to remap a sample tube that had already been remapped.
B31451AE
1
2
Left Drawer
Figure 1.7 Error
Lane) for the following reasons:
Note
If you manually load a sample into a sample tube carrier before the Error Lane, the
system sorts it to the Error Lane. If you manually load a sample into a sample
carrier past the Error Lane, the system moves the sample tube around the track
generating errors as it moves.
Inlet and Outlet Module Description
3. Universal Centrifuge Bypass Rack: WHITE
4. Universal Remap Rack: YELLOW
3
4
Dynamic Inlet Module
5
6
Right Drawer
1
7
8
1-5