Troubleshooting
Error Code Tables
Table 4.7
Outlet Error Codes (Continued)
Sensor
BUZZ 01-0*
FS021 ON
LS001 OFF
LS002 OFF
LS003 OFF
LS004 OFF
LS005 OFF
LS006 OFF
LS007 OFF
4-34
Problem
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
Inlet Rack #1 detection
error. Inlet Rack #1 is not
present or not detected by
the rack sensor.
Inlet Rack #3 detection
error. Inlet Rack #3 is not
present or not detected by
the rack sensor.
Inlet Rack #5 detection
error. Inlet Rack #5 is not
present or not detected by
the rack sensor.
Inlet Rack #7 detection
error. Inlet Rack #7 is not
present or not detected by
the rack sensor.
Inlet Rack #2 detection
error. Inlet Rack #2 is not
present or not detected by
the rack sensor.
Inlet Rack #4 detection
error. Inlet Rack #4 is not
present or not detected by
the rack sensor.
Inlet Rack #6 detection
error. Inlet Rack #6 is not
present or not detected by
the rack sensor.
1.
Restart
Select
on the control panel to retry.
2.
Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3.
If the error persists, select
to release the carrier.
1.
Inspect and clean the fiber optic sensor head.
2.
Restart
Select
on the control panel to retry.
3.
If the gripper still cannot grab the sample tube from the
carrier, remove the tube and close the doors.
4.
Restart
Select
to continue and then place the sample
tube In the correct outlet rack and position.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Solution