Beckman Coulter Power Express Instructions For Use Manual page 102

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.6
Bulk Inlet Error Codes (Continued)
Sensor
FS021 OFF
FS021 ON
FS025 OFF
FS131 OFF
FS131 ON
FS231 OFF
FS231 ON
4-26
Problem
At FS021 area, sample tube
is not detected.
Sample tube is detected at
FS021 area.
Tube not detected in
sample carrier at SN025.
At FS131 and PS131 areas,
sample tube is not
detected.
At FS131 and PS131 areas,
sample tube is detected
while the area should be
empty.
At FS231 and PS231 areas,
sample tube is not
detected.
At FS231 and PS231 areas,
sample tube is detected
while the area should be
empty.
1.
Confirm that the sample tube is placed in the sample
carrier on the track (FS021).
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sample tube is transferred to the next
process.
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sample tube is in the sample carrier at
SN025.
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that there is no obstruction to move the sample in
the PS131 area.
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that there is no obstruction in the FS131 and
PS131 areas.
2.
Remove tube by tilting tube orienter. Tilt tube orienter
back to home position (sensor light will be on).
3.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that there is no obstruction to move the sample in
the FS231 and PS231 areas.
2.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that there is no obstruction in the FS231 and
PS231 areas.
2.
Remove tube by tilting tube orienter. Tilt tube orienter
back to home position (sensor light will be on).
3.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE