Table 4.6
Bulk Inlet Error Codes (Continued)
Sensor
Passage sensor error. A
SN012 PAS
passage sensor does not
SN016 PAS
detect a carrier within a
predetermined timeout
period.
Sample tube is placed
SN017 ERR
upside down or the cap has
come off the sample tube.
Passage sensor error. A
SN022 PAS
passage sensor does not
SN026 PAS
detect a carrier within a
predetermined timeout
period.
Sample tube is placed
SN027 ERR
upside down or the cap has
come off the sample tube.
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
1.
Inspect for spilled sample tube, missing or loose cap, and
clean as necessary.
2.
Inspect the sample tube and confirm that it is acceptable
for further processing. Remove the tube and place it in an
appropriate rack at the Inlet module if the cap or tube
appears loose, damaged, or missing. Do not leave a sample
tube with a missing or loose cap in the Bulk Inlet module.
3.
Inspect for sample tube placed in sample carrier upside
down. Manually remove and replace tube right side up.
4.
Select
removed from position SN017. Select
continue otherwise.
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
When no carrier is found, it passes through this position
after the time limit.
3.
Select
operation.
1.
Inspect for spilled sample tube, missing or loose cap, and
clean as necessary.
2.
Inspect the sample tube and confirm that it is acceptable
for further processing. Remove the tube and place it in an
appropriate rack at the Inlet module if the cap or tube
appears loose, damaged, or missing. Do not leave a sample
tube with a missing or loose cap in the Bulk Inlet module.
3.
Inspect for sample tube placed in sample carrier upside
down. Manually remove and replace tube right side up.
4.
Select
removed from position SN027. Select
continue otherwise.
Solution
Restart
on the control panel to resume
Clear
to continue if the sample tube has been
Restart
on the control panel to resume
Clear
to continue if the sample tube has been
4
Troubleshooting
Error Code Tables
Restart
to
Restart
to
4-31