Inlet and Outlet Module Procedures
Clean the Bulk Inlet Module
Figure 2.7 Bulk Inlet Control Panel - Cleaning Mode
4
Remove the top interlock covers.
5
Clean the transfer belts, tube grippers and other parts or surfaces that have been
contaminated with 10% bleach or laboratory cleaning solution.
6
Push open the bin covers and clean the interior surfaces with 10% bleach or laboratory
cleaning solution.
7
Close the bin covers. The bin covers lock, and the feeders perform one operation cycle
to expose a different area of the interior surface of the bins.
8
Repeat steps 6 and 7 until the interior surfaces of the bins are cleaned.
9
Push to open the spill containment door and pull out the spill containment tray on the
lower front of the module. Remove and appropriately dispose of the customer-supplied
laboratory absorbent material and clean the tray with 10% bleach or laboratory
cleaning solution. Refer to
2-10
Caution
Before you clean the Bulk Inlet, wear appropriate gloves for biohazard protection.
Use caution when you handle the internal components of the module as they can
have sharp edges.
Figure 2.8 Spill Containment
Tray.
B31451AE