Beckman Coulter Power Express Instructions For Use Manual page 98

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.5
Bulk Inlet Error Codes
Sensor
DM132A ON
DM132B ON
DM132 OFF
DM141A ON
DM141B ON
DM141 OFF
DM201A ON
DM201B ON
DM201 OFF
DM205A ON
DM205B ON
DM205 OFF
4-22
Problem
Tube Orientation Detection
is not working properly. *A
ON Error; Upward motion
failed. *B ON Error;
Downward motion failed.
*OFF Error; Sensors on both
ends became ON.
Uncapped tube is detected.
Tube Supply Box Z-axis
Lifter is not working
properly. *A ON Error;
failed to close. *B ON Error;
failed to open. *OFF Error;
Sensors on both ends
became ON.
Tube Supply Box door is not
working properly. *A ON
Error; Upward motion
failed. *B ON Error;
Downward motion failed.
*OFF Error; Sensors on both
ends became ON.
1.
Inspect for and remove any obstructions.
2.
Confirm that the tube is fully inserted in the tube
orientation detector.
3.
If an uncapped tube is found, remove the tube and clean
the Bulk Inlet area.
4.
Confirm that the air compressor is on and the auto switch
is working correctly.
5.
Confirm that the compressed air supply is within the
normal pressure range.
6.
Rotate DM132 to its home position (horizontal) if
necessary.
7.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
If uncapped tube is found, remove tube and clean Bulk
Inlet area. Inspect the sample tube and confirm that it is
acceptable for further processing. Place the tube in the
appropriate rack in the Inlet Module. Do not replace the
sample tube in the Bulk Inlet Module.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the sensors are working correctly.
2.
Inspect for and remove any obstructions.
3.
Restart
Select
on the control panel to resume
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE