Safety Notice
General Warnings and Cautions
General Warnings and Cautions
Warning
Do not use the equipment in a manner not specified by Beckman Coulter, Inc., as the
protection provided by the equipment can be impaired.
Warning
Each sample tube processed by the system must have a unique sample ID and
readable bar code label. The system sends sample tubes with duplicate sample IDs to
the Error Lane. Damaged or unreadable bar code labels cause errors.
If you load multiple sample tubes with the same sample ID on the automation system
and on a connected analyzer at the same time, the system can send duplicate results
to the LIS. Contact your Beckman Coulter Representative for suggestions to
implement unique sample ID labeling.
Caution
To reduce risk of personal injury, operate the system only with all covers in place.
Caution
Do not load or view the Instructions for Use PDF files onto any computer connected
to the automation system. Failure to follow this caution can reduce computer
processing speed and system performance.
Caution
Use only the approved Power Express parts and supplies as listed in the Power
Express General System Operation IFU, Appendix B. Use only the approved sample
tubes as noted in the Power Express General System Operation IFU, Operational
Overview.
Electromagnetic Wave and Noise Precautions
This In Vitro diagnostic (IVD) equipment complies with the emission and immunity
requirement described in IEC 61326-2-6.
Caution
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment, it could cause radio interference, in which case, you might need to take
measures to mitigate the interference.