Beckman Coulter Power Express Instructions For Use Manual page 47

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Figure 2.6 Bin Feeder Filled and Ready to Start
6
When completed, inspect the bins for any unprocessed samples.
Clean the Bulk Inlet Module
The Bulk Inlet module is designed to handle sample tubes with the original manufacturer's
caps attached and the vacuum seal intact. If sample tubes with other types of caps are used,
or tubes have had the vacuum released between the tube and the cap, it is possible for a
sample tube to spill in the process of being loaded by the Bulk Inlet. Use this procedure to
clean the Bulk Inlet Module.
Warning
In the event of a spill, sample tubes that have already been loaded to the system may
be affected. Pause the system and inspect the sample tubes and modules for
contamination and clean as appropriate.
At the Bulk Inlet control panel:
1
Select Manual
2
Select Maintenance
3
Enter function code 31 and select Enter
is displayed on the control panel. Refer
Mode.
B31451AE
and OK
to place the module in Manual mode.
and select the number field to display the onscreen keyboard.
Figure 2.7 Bulk Inlet Control Panel - Cleaning
Inlet and Outlet Module Procedures
Clean the Bulk Inlet Module
twice to start the cleaning process. Cleaning
2
2-9