RM-A (Remap-Ambient Storage) or select it again to return it to being unselected (
Select Restart
When RM-A is selected, the user is prompted with RM-A Selected. Continue Run?
Select OK
Cancel
Selecting Inlet Module Priority
The Inlet and Bulk Inlet operate with a priority system. It is used to determine the order
that samples are loaded onto the system from the bins or racks of the Bulk Inlet and Inlet
modules. The default priority order is:
• Priority Rack: RED (ST)
• Routine Rack: BLUE (RT)
• Centrifuge Bypass Rack: WHITE (BP)
• Remap Rack: YELLOW (RM)
• Bin Feeder #1 (BULK-1)
• Bin Feeder #2 (BULK-2)
The user can configure these priorities to better suit their laboratory operation and
preferences.
At the Inlet control panel:
1
Select Manual
2
Select Maintenance
3
Enter function code 83 and select Enter
Refer to
Figure 2.3 Rack Priority Setting - Default Settings
B31451AE
when the desired remap destination is displayed.
to continue with loading samples with the Remap enabled for Ambient or
to change your selection.
and OK
to place the module in Manual mode.
and select the number field to display the onscreen keyboard.
Figure 2.3 Rack Priority Setting - Default
Inlet and Outlet Module Procedures
Selecting Inlet Module Priority
twice to display the priority setup screen.
Settings.
2
).
2-5