Procedures - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Introduction
This chapter provides information to identify and resolve errors related to the Inlet and
Outlet modules, as well as system operational errors that are not specific to any Power
Express module.
Note
For error recovery information for other Power Express modules, refer to the
appropriate Power Express module Instructions for Use.
Warning
Always operate the system with all shields and doors in position and secured to avoid
injury.
Perform system operations with caution.
Wear Personal Protective Equipment (PPE) such as gloves, eye shields, and lab coats.
Wash hands thoroughly after contact with sample media and all maintenance
activities.
Observe all laboratory policies and procedures related to the handling of
biohazardous materials.
Refer to the applicable sources (such as Material Safety Data Sheets) for specific
hazard information.
Caution
Before you perform an error recovery procedure, confirm the module is stopped.
When the module is stopped, the control panel turns red.
Caution
Even when the Inlet and the Outlet modules are in Stop or Pause mode, the air
system is still active and applies a constant air pressure to the gripper arm assembly.
This can cause unexpected movement of the gripper arm assembly and create a pinch
hazard when you resolve jammed object errors at these modules. Use caution when
you resolve jammed object errors.
To resolve operational issues not discussed in this chapter, contact a Beckman Coulter
representative.
B31451AE
Error Recovery Procedures
CHAPTER 3