3. Samples being loaded by the Storage module must be reloaded. Place the sample
tubes into spare sample holders and insert the sample holders at the T-lane Entry
point. Refer to the diagrams below.
Note
Figure 3.1 Sample Tube Enter Lane
Figure 3.2 Second Stockyard Enter Lane
B31451AE
Any samples removed from a Storage module with any floor configured for Sarstedt
tubes, the sample tube cannot be reloaded. You must remove the sample tube
from the track and store it offline.
Error Recovery Procedures
System Error Recovery Procedures
3
3-9