Table 4.6
Bulk Inlet Error Codes (Continued)
Sensor
Stopper error. The stopper
SL011 OFF
is not working correctly.
SL011 ON
ON error: The cylinder does
SL012 OFF
not respond correctly to an
SL012 ON
ON signal from a PLC.
SL015 OFF
SL015 ON
OFF error: The cylinder
SL016 OFF
does not respond correctly
to an OFF signal from a PLC.
SL016 ON
SL021 OFF
SL021 ON
SL022 OFF
SL022 ON
SL025 OFF
SL025 ON
SL026 OFF
SL026 ON
SL031 OFF
SL031 ON
SL032 OFF
SL032 ON
SL041 OFF
SL041 ON
SL042 OFF
SL042 ON
SL091 OFF
SL091 ON
SL092 ON
SL092 OFF
Tube clamp is not working
SL017 OFF
properly. *ON Error:
SL017 ON
Gripper was not able to
SL027 OFF
move to clamp. *OFF Error:
Gripper was not able to
SL027 ON
release.
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
1.
Confirm that the air compressor is on and the auto switch
is working correctly.
2.
Confirm that the compressed air supply is within the
normal pressure range.
3.
Confirm that the tube is positioned correctly and that there
are no obstructions.
4.
Select
operation.
Solution
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-29