Bulk Inlet Module Error Recovery Procedure; Outlet Module Error Recovery Procedure - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Bulk Inlet Module Error Recovery Procedure

Bulk Inlet module errors Inlet module errors can also include:
• Bin feeder errors
• Belt transfer errors
• Gripper finger errors
• Transfer-arm blockage
• Sensor malfunction
• Jammed sample tube carrier
Caution
Before you perform an error recovery procedure, confirm that the module is stopped.
When the module is stopped, the control panel turns red.
1
Select Mute
2
Read the error code and the description on the control panel.
3
Locate the error code in the
4
Resolve the condition causing the error message.
5
Select Restart

Outlet Module Error Recovery Procedure

Outlet module errors can include:
• Operator Rack Load error
• Bar code reader error
• Sensor malfunction
• Gripper finger error
• Transfer-arm blockage
• Jammed sample tube carrier
Caution
Before you perform an error recovery procedure, confirm the module is stopped.
When the module is stopped, the control panel turns red.
1
Select Mute
2
Read the error code and the description on the control panel.
B31451AE
on the control panel to silence the alarm.
Bulk Inlet Error Codes
on the control panel to resume operation.
on the control panel to silence the alarm.
Error Recovery Procedures
Inlet and Outlet Module Recovery Procedures
table in Chapter 4.
3
3-19