Beckman Coulter Power Express Instructions For Use Manual page 99

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.5
Bulk Inlet Error Codes (Continued)
Sensor
Tube Supply Box Z-axis
DM211A ON
Lifter is not working
DM211B ON
properly. *A ON Error;
DM211 OFF
failed to close. *B ON Error;
failed to open. *OFF Error;
Sensors on both ends
became ON.
Tube Supply Box door is not
DM215A ON
working properly. *A ON
DM215B ON
Error; Upward motion
DM215 OFF
failed. *B ON Error;
Downward motion failed.
*OFF Error; Sensors on both
ends became ON.
Left side tube transfer belt
DM221A ON
2 stopper 1 is not working
DM221B ON
properly. *A ON Error;
DM221 OFF
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
Left side tube transfer belt
DM223A ON
stopper 2 area is not
DM223B ON
working properly. *A ON
DM223 OFF
Error; Failed to move to the
supply box side. *B ON
Error; Failed to move to the
tube clamp holds side. *OFF
Error; Sensors on both ends
became ON.
Left side tube transfer belt
DM225A ON
stopper 3 is not working
DM225B ON
properly. *A ON Error;
DM225 OFF
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
Left side tube transfer belt
DM227A ON
stopper 4 is not working
DM227B ON
properly. *A ON Error;
DM227 OFF
Failed to extend. *B ON
Error; Failed to retract.
*OFF Error; Sensors on both
ends became ON.
B31451AE
Problem
1.
Confirm that the sensors are working correctly.
2.
Inspect for and remove any obstructions.
3.
Select
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Select
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Select
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Select
operation.
1.
Inspect for and remove any obstructions.
2.
Confirm that the sensors are working correctly.
3.
Select
operation.
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-23