Troubleshooting
Error Code Tables
Table 4.3
Error Lane Error Codes (Continued)
Sensor
BUZZ01-0*
BUZZ02- 0*
CA1 ERR
CA2 ERR
CA1PNG TIM
CA2PNG TIM
4-12
Problem
Line Control CPU Error, bar
code label, or RFID read
error or Sample Positioning
error.
Lane 1 Camera fails to read.
Lane 2 Camera fails to read.
CA1 Lane 1 Camera fails to
communicate.
CA2 Lane 2 Camera fails to
communicate.
1.
Restart
Select
on the control panel to retry.
2.
Confirm that the bar code reader or RFID reader reads the
bar code label or RFID tag correctly. If necessary, reprint
the bar code label and re-apply it to the sample tube.
3.
If the error persists, select
to release the carrier.
1.
Restart
Select
on the control panel to resume
operation.
2.
If problem continues, power off track to reset camera unit.
3.
If the error still occurs, disable lane 1 operation.
4.
If the error still occurs, contact your Beckman Coulter
representative.
1.
Restart
Select
on the control panel to resume
operation.
2.
If problem continues, power off track to reset camera unit.
3.
If thee error still occurs, disable lane 2 operation.
4.
If the error still occurs, contact your Beckman Coulter
representative.
1.
Restart
Select
on the control panel to resume
operation.
2.
If problem continues, power off track to reset camera unit.
3.
If problem still continues, disable lane 1 operation and
contact Beckman Coulter service representative.
1.
Restart
Select
on the control panel to resume
operation.
2.
If problem continues, power off track to reset camera unit.
3.
If problem still continues, disable lane 2 operation and
contact Beckman Coulter service representative.
Solution
Clear
on the control panel
B31451AE

Hide quick links:

Advertisement

Table of Contents
loading