Table 4.8
T Lane Error Codes (Continued)
Sensor
Established communication
CCIE ERR
between PLC and LC has
been disrupted.
Failed to send a message to
CCIE SND
the line controller.
PLC hardware error. A PLC
OP CALL
battery may have run out.
Passage sensor error. A
PS065 PAS
passage sensor does not
detect a carrier within a
predetermined timeout
period.
The PLC module (QJ71GF)
QJ71GF ERR
used for the
communication between
the PLC and LC
malfunctioned.
RFID initialization error.
RF01 INI
RF02 INI
RF03 INI
PLC register settings have
REG RST
been reset. They may have
reverted to their default
values.
Stopper error. The stopper
SL001 ON
is not working correctly.
SL001 OFF
ON error: The cylinder does
SL002 ON
not respond correctly to an
SL002 OFF
ON signal from a PLC.
SL003 ON
SL003 OFF
OFF error: The cylinder
SL004 ON
does not respond correctly
to an OFF signal from a PLC.
SL004 OFF
B31451AE
Problem
1.
Confirm the cable connection between LC and PLC, located
at the upper stream side of the Error unit.
2.
Confirm all the PLCs within the line are powered on.
3.
When the communication between PLC and LC has been
re-established, an error is recovered.
4.
Select
Function code 01 (Lane through (Bypass)) to continue.
Confirm that the cables between individual PLCs and the Line
Control are connected correctly.
Replace the battery if the battery indicator on the front of the
PLC box is blinking.
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
1.
Reset the PLC
2.
If the reset does not recover from the error, QJ71GF
module for the PLC can be defective.
1.
Confirm that the RFID cable is connected correctly.
2.
Reboot the module.
3.
Place the module in
Inspect and adjust the PLC register settings as required.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
Solution
Restart
on the control panel to retry or use
Restart
on the control panel to resume
Run
mode to resume operation.
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-41