Beckman Coulter Power Express Instructions For Use Manual page 49

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Figure 2.8 Spill Containment Tray
10
Insert a new piece of customer-supplied laboratory absorbent material in the bottom of
the spill containment tray and place the tray back in position. Close the spill
containment door.
11
Inspect and clean other parts of the Power Express system as necessary with 10%
bleach or laboratory cleaning solution.
12
Select Enter
13
Select Auto
If you are unable to completely clean the Bulk Inlet module, use the Inlet, Error Lane
and Bulk Inlet Enable and Disable procedure to disable the lane that you are unable to
clean.
Remove Sample Tubes from the Error Lane
When ten sample tubes accumulate in the Error Lane, the system sounds an alarm,
activates a flashing beacon, and pauses processing. Any additional sample tubes with errors
are queued on the track. Use this procedure to remove the tubes from the Error Lane.
If configured, the flashing beacon activates each time a sample goes to the Error lane. (V3.0
or later)
B31451AE
and Exit
to finish the cleaning process.
and OK
to resume operation.
Inlet and Outlet Module Procedures
Remove Sample Tubes from the Error Lane