Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

2
Reset all Stop buttons that were pressed by pressing the buttons again. You cannot
restore power to the system until all Stop buttons have been reset.
3
Press the green ON button on the Inlet module to restore power to the system. Resolve
the conditions that prompted the emergency stop. The modules do not yet home or
process any sample tubes.
4
At the Inlet module, remove any tubes from the grippers and return them to the
appropriate Inlet rack:
a. If a stopper is jamming a sample tube carrier, push it forward to release the carrier.
b. If the transfer arm is extended, follow the procedure below to retract the arm to the
standby position.
— Select Manual
— Select Maintenance
— Select the number field to display the onscreen keyboard.
— Enter function code 06 and select Enter
c. If there are sample tubes in the sample tube carriers in the tube loading area,
remove the tubes from the carriers.
d. Remove sample tubes in the sample tube carriers at the RF01 and RF02 positions.
e. Determine which rack or Bulk Inlet the samples removed in steps c and d were
loaded from. Reload the sample tubes to racks in the Inlet module or the Bulk Inlet
bin.
5
At the Error Lane, if a stopper is jamming the sample tube carrier, push it forward to
release the carrier.
6
At the Bulk Inlet module, remove sample tubes from the transfer mechanism if a
stopper is jamming the sample carrier.
a. If a tube is at FS011 or FS021, and it is not fully seated in the carrier, push the
sample tube down
b. If a tube is at FS131 or FS231, remove the tube and replace it in the loader bin.
Rotate DM131 or DM231 to remove the tube and return the motor to a horizontal
position.
c. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
7
At the Outlet module, remove sample tube from the gripper.
a. If a sample tube is present in the gripper finger, manually remove the tube and
place it back into the empty sample tube carrier at BR01.
b. If the transfer arm is extended, follow the procedure below to retract the arm to the
standby position.
— Select Maintenance
— Select the number field to display the onscreen keyboard.
— Enter function code 06 and select Enter
B31451AE
and OK
to place the module in Manual mode.
on the control panel.
on the control panel.
Error Recovery Procedures
System Error Recovery Procedures
twice.
twice.