Beckman Coulter Power Express Instructions For Use Manual page 67

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

b. Perform this procedure from the Direct Track Sampling module control panel.
Select Maintenance
c. Select the number field to display the onscreen keyboard.
d. Enter function code 91 and select Enter
BR03 and to release tubes after bar code reader BR03. This ensures that the system
reads and registers the bar code IDs at BR03.
e. As the samples pass through the DTS sampling position make a note of the sample
ID numbers for later retrieval. See Note below.
f.
After the system releases all tubes from BR03 and BR02, select Enter
g. Select Auto
h. Perform troubleshooting on the connected analyzer as necessary to re-establish
communication with the Direct Track Sampling module.
i.
Use the Cennexus retrieval function to manually collect the samples, or use the LIS
console to order retests on the samples noted in step e above.
20
If the Abbott i2000SR Direct Track Sampling module T-Lane generates a CCIE error
when you press the main track Stop button:
a. Select Mute
b. Select Restart
21
At the T-Lanes, if a stopper is jamming the sample tube carrier, push it forward to
release the carrier. Leave other tubes and sample tube carriers alone.
22
At the Aliquotter and Labeler components:
Note
B31451AE
on the control panel.
and OK
Note
The analyzer does not aspirate samples exiting the DTS connection. If the
analyzer is a dynamic analyzer (AU680, DxC 700 AU, or DxI), Cennexus
automatically recognizes that the analyzer has not aspirated the samples and
reroutes them to another analyzer for testing. If the analyzer is a non-dynamic
analyzer, the system considers aspiration to be complete, and you must
manually request retesting or retrieve samples using Cennexus for offline
testing.
and Clear
on the control panel to resume operation.
Note
Pressing the main track Stop button does not affect the Abbott i2000SR,
because the Abbott i2000SR is on a separate power system, but doing so does
generate the CCIE error.
Regardless of the status of the Aliquotter and Labeler, the recovery procedure is
the same.
twice to hold tubes at bar code reader
to resume operation.
on the control panel to clear the errors.
Error Recovery Procedures
System Error Recovery Procedures
3
and Exit
.
3-13