Outlet Error Codes - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.6
Bulk Inlet Error Codes (Continued)
Sensor
The orientation of the
SN241 OFF
sample tube is not correctly
SN241 ERR
detected.
Left supply bin cover is
SN251 OFF
open.
SN252 area detects
SN252 ON
obstacles.

Outlet Error Codes

Table 4.7
Outlet Error Codes
Sensor
Bar code reader
BR01 INI
initialization error.
Tube Held
B31451AE
Problem
1.
Confirm that the sample tube is fully positioned in the tilt
sensor.
2.
Confirm that the sample tube is capped. If an uncapped
tube is found, remove the tube and clean the Bulk Inlet
area and any sample tubes that can be affected. Inspect
the sample tube and confirm that it is acceptable for
further processing. Place the tube in the appropriate rack
in the Inlet Module. Do not replace the sample tube in the
Bulk Inlet Module.
3.
Select
1.
Manually close the left supply bin cover.
2.
Select
operation.
1.
Confirm that there is no obstacle in the SN252 area.
2.
If so, remove it manually.
3.
Select
operation.
Problem
1.
Confirm that the bar code reader cable is connected
correctly.
2.
Reboot the module.
3.
Place the module in
Occurs after a Stop Button or manual reset of the Outlet
module when the gripper is holding a sample tube. Refer to
Tube Held Recovery
Solution
Restart
to resume operation or select
to cancel the process.
Restart
on the control panel to resume
Restart
on the control panel to resume
Solution
Run
mode to resume operation.
Procedure.
4
Troubleshooting
Error Code Tables
Cancel
4-33