Beckman Coulter Power Express Instructions For Use Manual page 84

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.2
Inlet Error Codes (Continued)
Sensor
SL006 ON
SL006 OFF
SL007 ON
SL007 OFF
4-8
Problem
Locking cylinder error. The
cylinder that locks the left
drawer lock is not working
correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Locking cylinder error. The
cylinder that locks the right
drawer lock is not working
correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1.
Confirm that the drawer is fully closed.
2.
Confirm that the cylinder is getting enough air.
3.
Confirm that the cylinder motion sensor is working
correctly.
1.
Confirm that the drawer is fully closed.
2.
Confirm that the cylinder is getting enough air.
3.
Confirm that the cylinder motion sensor is working
correctly.
Solution
B31451AE