Troubleshooting
Error Code Tables
Table 4.3
Error Lane Error Codes (Continued)
Sensor
SL031 ON
SL031 OFF
SL032 ON
SL032 OFF
SL033 ON
SL033 OFF
SL034 ON
SL034 OFF
SL035 ON
SL035 OFF
SL041 ON
SL041 OFF
SL042 ON
SL042 OFF
SL043 ON
SL043 OFF
SL044 ON
SL044 OFF
4-16
Problem
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Diverter error. The diverter
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Diverter error. The diverter
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents