23
At the Secondary Decapper module:
a. If there is a sample tube at BR01, note the Sample ID for use in step c before
continuing with the following recovery steps.
1. If a stopper is jamming the sample tube carrier, push it forward to release the
2. Check for any sample tube that is at the decapping position – Will cause an
3. Leave other tubes in carriers.
b. Place the Secondary Decapper module into Auto mode. Look for any dropped caps
and discard them down the cap chute.
c. Place the sample tube removed in a.2.1 above into the empty sample carrier as it
exits the module.
d. If the error PS041, PS042 or PS043 is generated for the sample tube that was at
BR01 (Confirm the sample ID noted in step a.), manually remove the cap.
e. Select Clear
release the sample tube.
f.
Select Clear
g. Select Auto
24
At the Decapper module:
a. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
b. Check for any tube that is ready to be decapped at the decapping positions SN006
or SN016. Remove these tubes, manually decap, and discard the caps in the cap
chute and place the tube back in the sample carrier.
c. Select Auto
d. If the error PS011, PS012, PS111, PS112 occurs, select Clear
panel to clear the error and release the sample tube.
e. If PS014 or PS114 error is generated, manually decap the tube and dispose of it in
the cap chute.
f.
Select Clear
the sample tube.
25
At the Level Detector component:
B31451AE
b. On the Labeler control panel, select Enter
c. Select Auto
and OK
processing.
carrier.
PS044 ON error.
1. Remove the sample tube.
2. Using Cennexus, determine if the sample tube is supposed to be decapped
or not.
3. If it is supposed to be decapped, manually remove the cap and dispose of
the cap down the cap chute.
from the control panel to clear the PS041, PS042 or PS043 and
or Restart
and OK
and OK
from the control panel to clear the PS014 or PS114 error and release
on both the Aliquotter and Labeler to resume
from the control panel to clear any other errors.
to resume operation.
to resume operation.
Error Recovery Procedures
System Error Recovery Procedures
and Exit
.
from the control