Beckman Coulter Power Express Instructions For Use Manual page 120

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.8
T Lane Error Codes (Continued)
Sensor
SL035A ON
SL035B ON
SL035 OFF
SL036A ON
SL036B ON
SL036 OFF
SL041 ON
SL041 OFF
SL042 ON
SL042 OFF
SL043 ON
SL043 OFF
SL044 ON
SL044 OFF
SL045A ON
SL045B ON
SL045 OFF
SL046A ON
SL046B ON
SL046 OFF
SL051 ON
SL051 OFF
SL052 ON
SL052 OFF
SL053 ON
SL053 OFF
SL054 ON
SL054 OFF
4-44
Problem
Diverter error. The diverter
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Diverter error. The diverter
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE