Beckman Coulter Power Express Instructions For Use Manual page 86

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Troubleshooting
Error Code Tables
Table 4.2
Inlet Error Codes (Continued)
Sensor
SL051 ON
SL051 OFF
SL061 ON
SL061 OFF
SL062 ON
SL062 OFF
SL063 ON
SL063 OFF
SL071 ON
SL071 OFF
SL072 ON
SL072 OFF
4-10
Problem
The rear door locking
cylinder is not working
correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Transfer arm Z-axis motion
error.
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1.
Confirm that the rear doors are completely closed.
2.
Confirm that there are no obstructions around the rear
door locks.
3.
Confirm that the air compressor is on and the auto switch
is working correctly.
4.
Confirm that the compressed air supply is within the
normal pressure range.
5.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that the tube is seated correctly in the sample
carrier.
2.
Inspect for and remove any obstructions.
3.
Confirm that the air compressor is on and the auto switch
is working correctly.
4.
Confirm that the compressed air supply is within the
normal pressure range.
5.
Restart
Select
on the control panel to resume
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Restart
Select
on the control panel to resume
operation.
Solution
B31451AE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents