Enable or Disable Lanes at the Inlet, Error Lane, or Bulk Inlet
The Inlet, Error Lane, and Bulk Inlet all operate with a common two-lane sample carrier
transfer system. Each also operates with their own separate transfer mechanisms
connected to those two lanes.
Each module has a separate enable/disable mechanism used to isolate any specific error
conditions within, without interfering with the adjoining modules.
Inlet – Enables or disables the front or rear sample transfer points. This also affects the
Bulk Inlet and Error Lane operations.
Bulk Inlet – Enables or disables the left or right loading bins and front or rear sample
transfer points. This does not affect the Inlet or Error Lane operations.
Error Lane – Enables or disables the front or rear bar code readers, tube size detection, and
the front or rear Tube Type Detection system. This also affects the Inlet and Bulk Inlet
operations.
The user can Enable or Disable the front or rear lane operation of these modules using the
following procedure.
At the Inlet, Error Lane or Bulk Inlet control panels:
1
Select Manual
2
Select Maintenance
3
Enter function code 97 and select Enter
parameter followed by Enter
0 - Enable both lanes.
1 - Disable the front lane.
2 - Disable the rear lane.
3 - Disable both lanes, but leave the track functions operational.
4
After selecting the enable/disable parameter, select Enter
5
Select Auto
The enable/disable condition takes effect immediately. If entered at the Error Lane, the
corresponding Enable/Disable condition takes effect on the Inlet and Bulk Inlet. If
entered at the Inlet, the corresponding Enable/Disable condition takes effect on the
Bulk Inlet.
B31451AE
Enable or Disable Lanes at the Inlet, Error Lane, or Bulk Inlet
and OK
to place the module in Manual mode.
and select the number field to display the onscreen keyboard.
and OK
to resume operation.
Inlet and Outlet Module Procedures
twice. Then select the enable/disable
and Exit
2
.
2-7