Tube Held Recovery Procedure - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Error Recovery Procedures
Inlet and Outlet Module Recovery Procedures
3
Locate the error code in the
4
Resolve the condition causing the error message.
5
Select Restart

Tube Held Recovery Procedure

The Tube Held error occurs after a Stop Button or manual reset of the Outlet module, and
the gripper is holding a sample tube. The tube must be correctly removed and recovered
for the system to properly sort and track its status and allow the sample to be remapped
later.
1
Select Mute
2
Select Clear
3
Open the rear safety door of the Outlet module and remove the sample tube from the
gripper fingers. The gripper may be gently pulled toward the rear for easier access.
4
Determine the status of the sample:
If the sample needs to be remapped later or the
sample history tracking is important,
If the sample is removed from the system and
not remapped later,
5
Select Auto
The Outlet has two possible entry points for a sample to be re-introduced for error
recovery, depending on the configuration of your system. The Outlet is on a T-Lane with
a Storage module or without a Storage module.
3-20
Outlet Error Codes
on the control panel to resume operation.
on the Outlet control panel to silence the alarm.
and OK
.
If . . .
and OK
to resume processing.
table in Chapter 4.
Then . . .
Place the tube in an empty sample carrier and
place the sample carrier at the track entry
position nearest to the Outlet. Refer to
3.3 Outlet Sample Re-entry with or without a
Storage Module
.
Place the tube in the outlet rack below the
gripper. The sample is sorted; however, the
sample history and tracking information are
incomplete. Remapping of the sample is not
available.
Figure
B31451AE