Inlet and Outlet Module Description
Dynamic Inlet Module
• You tried to remap a sample tube that still has a related sample tube on the system.
• The system has previously read the same sample ID and it is still on the system.
• Processing for the sample tube requires modules that are not available, for example,
the Decapper or Aliquot module.
• The tube size is inconsistent with the required processing.
• The system cannot detect tube size or type of the sample tube.
The system routes up to ten tubes to the Error Lane before it requires operator
intervention. When ten sample tubes accumulate in the Error Lane, the system activates a
flashing beacon and an audible alarm.
For V3.0 or later, the Error lane provides an additional flashing beacon warning when a
sample is placed into the Error Lane. This warning may be configured or disabled. Contact
your Beckman Coulter Field Service Engineer to configure the Error Lane alert.
Figure 1.7 Error Lane
1. Safety Door
2. Stop Button
1-6
3. Control Panel
B31451AE