Type; Racks - Beckman Coulter Power Express Instructions For Use Manual

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Inlet and Outlet Module Description
Dynamic Inlet Module
Inlet Rack Type
The Inlet module uses four rack types, which you can place in any position in the rack
drawer, and use in any combination. Power Express processes the racks in a user defined
(V3.5 or later) priority order, independent of their physical location in the Inlet drawer.
Racks have the following default order:
• Priority Rack: RED
• Routine Rack: BLUE
• Centrifuge Bypass Rack: WHITE
• Remap Rack: YELLOW
There are two types of Inlet racks:
• Standard Racks
• Universal Racks
The Standard Racks are used when the system is running only 13 mm diameter tubes (13
mm x 75 mm, 13 mm x 90 mm, and 13 mm x 100 mm). The Universal Racks are used when
the system handles 15 mm or 16 mm diameter tubes or 15 mm/16 mm tubes intermixed
with 13 mm tubes. If 15 mm or 16 mm tubes are used, ensure that the racks placed in the
Inlet module are the Universal racks.
Figure 1.5 Inlet Standard Racks
1. Priority Rack
2. Routine Rack
1-4
3. Centrifuge Bypass Rack
4. Remap Rack
B31451AE