Beckman Coulter Power Express Instructions For Use Manual page 60

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Error Recovery Procedures
System Error Recovery Procedures
1
Select Manual
2
Select Maintenance
3
Select Bypass
4
Wait for all sample tube carriers to pass through the module.
5
Select Enter
6
Select Auto
Stop Button Recovery
When you press any yellow or white Stop button (except the Centrifuge Stop button), at any
time during operation, the system disconnects power to the Power Express system. The
Line Control and Cennexus remain ON. The air system also remains on.
Note
If you press the Centrifuge Stop button, refer to the Centrifuge Stop Button Recovery
Procedure, in the
Important
Press the Stop button to halt any hazardous condition.
Important
When you press the Stop button, the Line Control cannot accurately represent the
location of the last sample tube sorted by the system.
Caution
If you have an Aliquot module, when you press a Stop button, aliquot tips containing
sample drip very slowly. You may need to clean the drip trays. For more information,
refer to CHAPTER 5, Daily Maintenance Procedures, in the
IFU, PN
Important
Perform only the Stop button recovery steps that apply to your Power Express
configuration.
1
Do not stop the connected analyzers. The analyzers continue to process sample tubes.
3-6
and OK
to place the module in Manual mode.
on the control panel.
.
and select Exit
, then select OK
Centrifuge Module IFU, PN
B31448.
.
to resume processing.
B31456.
General System Operation