Beckman Coulter Power Express Instructions For Use Manual page 62

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Error Recovery Procedures
System Error Recovery Procedures
c. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
8
At each Storage module, remove sample tubes from the Storage module grippers:
a. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
b. Check the internal gripper fingers for a sample tube.
c. If there is a sample tube in the gripper, use the following procedure to remove the
tube:
d. If the loading and retrieval arm grippers and the shuttle cups are holding a tube,
remove it as follows:
9
Confirm the status in Cennexus or the LIS for sample tubes recovered from the loading
arm, retrieval arm, floor grippers, and shuttle cups to determine if the samples require
further processing (samples going into storage vs being retrieved).
1. Samples being retrieved by the Storage module must be processed manually offline.
When completed, store the tubes offline. They cannot be re-entered into the system.
2. Samples being retrieved for disposal must be manually disposed. Follow your
laboratory procedures for sample tube disposal.
3-8
— Select Maintenance
— Select the number field to display the onscreen keyboard.
— Enter function code 30 and select Enter
— Remove the tube. Check each shelf of the Storage module.
— Select Enter
and Exit
— Perform this procedure from the Storage control panel on the conveyor side.
Select Maintenance
— Select the number field to display the onscreen keyboard.
— Enter function code 32 and select Enter
— Select and hold Repeat
location.
— Remove sample tubes from Loading and Retrieval Arm grippers.
— Use a flexible claw-like grabber tool to reach inside the module to retrieve
sample tubes from the loading and unloading shuttle cups.
— Select Enter
and Exit
— Select Auto
and OK
Note
If any tube access doors were open when the Stop button was pressed, error
codes for those doors SLx21 or SLx23 are generated, where x represents the
number of the Storage module floor (1, 2, 3, or 4). Select Clear
the initialization process.
on the front control panel of the Storage module.
twice.
.
.
twice.
and manually move both arms to an accessible
.
.
to continue
B31451AE