Figure 3.3 Outlet Sample Re-entry with or without a Storage Module
1. Entry point for sample carrier when
T-Lane Module Error Recovery Procedure
T-Lane module errors can include:
• Diverter blockage
• Sensor malfunction
• Jammed sample tube carrier
Caution
Before you perform an error recovery procedure, confirm that the module is stopped.
When the module is stopped, the control panel turns red.
1
Select Mute
2
Read the error code and the description on the control panel.
3
Locate the error code in the
4
Resolve the condition causing the error message.
5
Select Restart
B31451AE
Outlet and Storage Modules share a
T-Lane
on the control panel to silence the alarm.
T-Lane Error
on the control panel to resume operation.
Error Recovery Procedures
Inlet and Outlet Module Recovery Procedures
2. Entry point for sample carrier when
Outlet is on its own T-Lane
Codes.
3
3-21