Beckman Coulter Power Express Instructions For Use Manual page 115

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.7
Outlet Error Codes (Continued)
Sensor
Left side drawer closure
SN001 OFF
error.
Right side drawer closure
SN002 OFF
error.
Passage sensor error. A
SN022 PAS
passage sensor does not
SN023 PAS
detect a carrier within a
predetermined timeout
period.
Module was reset while
Tube Held
gripper was holding a
sample tube.
B31451AE
Problem
1.
Confirm that the left side drawer is fully closed.
2.
Confirm that SN001 works correctly.
1.
Confirm that the right side drawer is fully closed.
2.
Confirm that SN002 works correctly.
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
3.
If no sample tube carrier is there or stopped, the passage
of the carrier did not trigger the sensor. Use a metal object
to trigger the sensor manually. Select
Occurs after a Stop Button or manual reset of the Outlet
module when the gripper is holding a sample tube. Refer to
Tube Held Recovery
Solution
Restart
on the control panel to resume
Procedure.
4
Troubleshooting
Error Code Tables
Restart
.
4-39