Beckman Coulter Power Express Instructions For Use Manual page 111

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.7
Outlet Error Codes (Continued)
Sensor
Inlet Rack #8 detection
LS008 OFF
error. Inlet Rack #8 is not
present or not detected by
the rack sensor.
Network protocol
NETC1 DATA
converter: command
execution error.
Network protocol
NETC1 RST1
converter: reminder to
NETC1 RST2
reboot the module. After
replacing a network
protocol converter and
configuring its settings for
the first time, the module
must be rebooted for the
settings to take effect.
Home sensor detection
PM001 JOG
error on the arm.
Loading arm positioning
PM001 POS
error.
Error generated while the
PM001 HOM
arm was moving to the
home position.
B31451AE
Problem
Confirm that the rack is there and set correctly in tray.
Clear the error to retry the command. If the error persists,
reboot the module.
Reboot the module.
1.
Confirm the X-axis arm motor moves towards the home
position and the home sensor works correctly.
2.
Place the module in
3.
Confirm that nothing interferes with the movement of the
loading arm.
4.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
Troubleshooting
Error Code Tables
Solution
Manual
mode.
Auto
mode to resume operation.
Manual
mode.
Auto
mode to resume operation.
Manual
mode.
Auto
mode to resume operation.
4
4-35