Table 4.8
T Lane Error Codes (Continued)
Sensor
Diverter error. The diverter
SL017 ON
is not working correctly.
SL017 OFF
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
SL021 ON
is not working correctly.
SL021 OFF
ON error: The cylinder does
SL022 ON
not respond correctly to an
SL022 OFF
ON signal from a PLC.
SL023 ON
SL023 0FF
OFF error: The cylinder
SL024 ON
does not respond correctly
to an OFF signal from a PLC.
SL024 OFF
Diverter error. The diverter
SL025A ON
is not working correctly.
SL025B ON
ON error: The cylinder does
SL025 OFF
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
SL026A ON
is not working correctly.
SL026B ON
ON error: The cylinder does
SL026 OFF
not respond correctly to an
SL031 ON
ON signal from a PLC.
SL031 OFF
SL032 ON
OFF error: The cylinder
SL032 OFF
does not respond correctly
to an OFF signal from a PLC.
SL033 ON
SL033 OFF
SL034 ON
SL034 OFF
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume