Beckman Coulter Power Express Instructions For Use Manual page 79

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.2
Inlet Error Codes (Continued)
Sensor
Established communication
CCIE ERR ERR
between PLC and LC has
been disrupted.
Failed to send a message to
CCIE SND
the line controller.
Inlet Rack #1 detection
LS001 OFF
error. Inlet Rack #1 is not
present or not detected by
the rack sensor.
Inlet Rack #3 detection
LS002 OFF
error. Inlet Rack #3 is not
present or not detected by
the rack sensor.
Inlet Rack #5 detection
LS003 OFF
error. Inlet Rack #5 is not
present or not detected by
the rack sensor.
Inlet Rack #7 detection
LS004 OFF
error. Inlet Rack #7 is not
present or not detected by
the rack sensor.
Inlet Rack #2 detection
LS005 OFF
error. Inlet Rack #2 is not
present or not detected by
the rack sensor.
Inlet Rack #4 detection
LS006 OFF
error. Inlet Rack #4 is not
present or not detected by
the rack sensor.
Inlet Rack #6 detection
LS007 OFF
error. Inlet Rack #6 is not
present or not detected by
the rack sensor.
Inlet Rack #8 detection
LS008 OFF
error. Inlet Rack #8 is not
present or not detected by
the rack sensor.
B31451AE
Problem
1.
Confirm the cable connection between LC and PLC, located
at the upper stream side of the Error unit.
2.
Confirm all the PLCs within the line are powered on.
3.
When the communication between PLC and LC has been
re-established, an error is recovered.
4.
Select
Function code 01 (Lane through (Bypass)) to continue.
Confirm the cables between individual PLCs and the Line
Controller are connected correctly.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Confirm that the rack is there and set correctly in tray.
Solution
Restart
on the control panel to retry or use
4
Troubleshooting
Error Code Tables
4-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents