Troubleshooting
Error Code Tables
Table 4.2
Inlet Error Codes (Continued)
Sensor
LS011 OFF
NETC1 DATA
NETC1 RST1
NETC1 RST2
No Carrier
OP.CALL
PM001 JOG
PM001 POS
PM001 HOM
4-4
Problem
Rack table detection error.
Network protocol
converter: command
execution error.
Network protocol
converter: reminder to
reboot the module. After
replacing a network
protocol converter and
configuring its settings for
the first time, the module
must be rebooted for the
settings to take effect.
Bulk Inlet is not providing
sample carriers to the Inlet.
PLC hardware error. A PLC
battery may have run out.
Home sensor detection
error on the arm.
Loading arm positioning
error.
Error generated while the
arm was moving to the
home position.
1.
Confirm that the rack table is set correctly.
2.
Confirm that LS011 works correctly.
Clear the error to retry the command. If the error persists,
reboot the module.
Reboot the module.
Confirm that the Bulk Inlet is in Run mode and that there are no
obstructions in the track lanes between the Bulk Inlet and the
Inlet.
Replace the battery if the BAT LED indicator on the front of the
PLC box is blinking.
1.
Confirm the X-axis arm motor moves towards the home
position and the home sensor works correctly.
2.
Place the module in
3.
Confirm that nothing interferes with the movement of the
loading arm.
4.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
Solution
Manual
mode.
Auto
mode to resume operation.
Manual
mode.
Auto
mode to resume operation.
Manual
mode.
Auto
mode to resume operation.