Beckman Coulter Power Express Instructions For Use Manual page 85

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.2
Inlet Error Codes (Continued)
Sensor
Stopper error. The stopper
SL011 ON
is not working correctly.
SL011 OFF
ON error: The cylinder does
SL012 ON
not respond correctly to an
SL012 OFF
ON signal from a PLC.
SL013 ON
SL013 OFF
OFF error: The cylinder
SL014 ON
does not respond correctly
to an OFF signal from a PLC.
SL014 OFF
SL015 ON
SL015 OFF
SL016 ON
SL016 OFF
SL021 ON
SL021 OFF
SL022 ON
SL022 OFF
SL023 ON
SL023 OFF
SL024 ON
SL024 OFF
SL025 ON
SL025 OFF
SL026 ON
SL026 OFF
SL031 ON
SL031 OFF
SL032 ON
SL032 OFF
SL041 ON
SL041 OFF
SL042 ON
SL042 OFF
Diverter error. The diverter
SL043 ON
is not working correctly.
SL043 OFF
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-9