Table 4.2
Inlet Error Codes (Continued)
Sensor
Transfer arm Z-axis motion
SL073 ON
error.
SL073 OFF
Left or right drawer closure
SN001 OFF
error.
SN002 OFF
Passage sensor error. A
SN012 PAS
passage sensor does not
SN013 PAS
detect a carrier within a
SN014 PAS
predetermined timeout
period.
SN022 PAS
SN023 PAS
SN024 PAS
SN032 PAS
SN043 PAS
SN0045 PAS

Error Lane Error Codes

Table 4.3
Error Lane Error Codes
Sensor
Bar code reader
BR01 INI
initialization error.
BR02 INI
B31451AE
Problem
1.
Confirm that the tube is seated correctly in the sample
carrier.
2.
Inspect for and remove any obstructions.
3.
Confirm that the air compressor is on and the auto switch
is working correctly.
4.
Confirm that the compressed air supply is within the
normal pressure range.
5.
Select
operation.
1.
Confirm that the drawers are closed correctly.
2.
Inspect for any obstructions in the drawer operation.
3.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
3.
If no sample tube carrier is there or stopped, the passage
of the carrier did not trigger the sensor. Use a metal object
to trigger the sensor manually. Select
Problem
1.
Confirm that the bar code reader cable is connected
correctly.
2.
Reboot the module.
3.
Place the module in
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
Solution
Run
mode to resume operation.
4
Troubleshooting
Error Code Tables
Restart
.
4-11