Beckman Coulter Power Express Instructions For Use Manual page 121

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.8
T Lane Error Codes (Continued)
Sensor
Diverter error. The diverter
SL055A ON
is not working correctly.
SL055B ON
ON error: The cylinder does
SL055 OFF
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
Stopper error. The stopper
SL056A ON
is not working correctly.
SL056B ON
ON error: The cylinder does
SL056B OFF
not respond correctly to an
SL061 ON
ON signal from a PLC.
SL061 OFF
SL062 ON
OFF error: The cylinder
SL062 OFF
does not respond correctly
to an OFF signal from a PLC.
SL063 ON
SL063 OFF
SL064 ON
SL064 OFF
Diverter error. The diverter
SL065A ON
is not working correctly.
SL065B ON
ON error: The cylinder does
SL065 OFF
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
1.
Confirm that a carrier is not jammed or stuck.
2.
Confirm that the air compressor is on and the auto switch
is working correctly.
3.
Confirm that the compressed air supply is within the
normal pressure range.
4.
Select
operation.
Solution
Restart
on the control panel to resume
Restart
on the control panel to resume
Restart
on the control panel to resume
4
Troubleshooting
Error Code Tables
4-45