Table 4.8
T Lane Error Codes (Continued)
Sensor
Passage sensor error. A
SN002 PAS
passage sensor does not
SN005 PAS
detect a carrier within a
SN007 PAS
predetermined timeout
period.
SN009 PAS
SN013 PAS
SN015 PAS
SN022 PAS
SN025 PAS
SN027 PAS
SN029 PAS
SN032 PAS
SN035 PAS
SN037 PAS
SN039 PAS
SN042 PAS
SN045 PAS
SN047 PAS
SN049 PAS
SN052 PAS
SN055 PAS
SN057 PAS
SN059 PAS
SN062 PAS
SN065 PAS
SN067 PAS
SN068 PAS
SN069 PAS
SN072 PAS
SN402 PAS
SN404 PAS

Sensor Diagrams

Use the sensor diagrams to locate errors on the system, in conjunction with Error Code
tables.
Background Information
Each hardware module uses a series of magnetic, fiber optic and other sensors to detect
sample tube carriers, sample tubes in sample tube carriers, hardware positioning, and
whether or not a gripper has grasped a tube.
B31451AE
Problem
1.
Confirm that a carrier is not jammed or stuck and that the
belt moves.
2.
Select
operation.
3.
If no sample tube carrier is there or stopped, the passage
of the carrier did not trigger the sensor. Use a metal object
to trigger the sensor manually. Select
Solution
Restart
on the control panel to resume
4
Troubleshooting
Sensor Diagrams
Restart
.
4-47