Beckman Coulter Power Express Instructions For Use Manual page 81

Inlets and outlets modules for in vitro diagnostic use
Hide thumbs Also See for Power Express:
Table of Contents

Advertisement

Table 4.2
Inlet Error Codes (Continued)
Sensor
Motor error.
PM001 ERR
The current register setting
PM001 LIM
forced the arm to move
beyond the allowable limit
position.
Home sensor detection
PM002 JOG
error on the arm.
Loading arm positioning
PM002 POS
error.
Error generated while the
PM002 HOM
arm was moving to the
home position.
B31451AE
Problem
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
1.
Place the module in
2.
Place the module in
3.
If the error still occurs, contact your Beckman Coulter
representative.
1.
Confirm the Y-axis arm motor moves towards the home
position and the home sensor works correctly.
2.
Place the module in
3.
Confirm that nothing interferes with the movement of the
loading arm.
4.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
1.
Place the module in
2.
Confirm that nothing interferes with the movement of the
loading arm.
3.
Place the module in
Solution
Manual
Auto
mode to resume operation.
Note
If the error still occurs, reboot the module, then
place it in Auto mode to resume operation.
Manual
Auto
mode to resume operation.
Manual
Auto
mode to resume operation.
Manual
Auto
mode to resume operation.
Manual
Auto
mode to resume operation.
Troubleshooting
Error Code Tables
mode.
mode.
mode.
mode.
mode.
4
4-5